The IOS LUKS-1550-TWM Laser Ultrasonic Kit for
Starters is designed to provide all components
necessary for establishing laser ultrasonic
inspection capability for laboratory use. It
includes the innovative AIR-1550-TWM laser
ultrasonic receiver, having an ideal combination of
sensitivity and response time and operating at the
eyesafe wavelength of 1550 nm. The fiber-coupled
measurement head is small and reconfigurable. With
added integration, the LUKS-1550-TWM can perform
process monitoring and inspection in factory
conditions and in-service inspection in the field.
Industries Served:
• Aerospace
• Auto
• Oil and gas pipeline
• Steel and cast iron
• Shipbuilding
• Semiconductor
• Electronic components and
packages
• Medical devices
• Ceramics
• Glass bottling
|
|
|
Materials Inspected:
•
Steel
•
Cast iron
•
Ceramics
•
Glass
•
Composites
•
Semiconductors |
|
Typical Applications:
• Thickness Measurements
• Glass containers
• Ceramic and metallic
coatings
• Steel tubes and cast iron
pipes
• Defect Detection
• Laser welds
• Ceramic coatings
• Electronic packages
• Adhesives
• Small parts |
|
The receiver and probe laser are mounted in a 19” rack.
The remote measurement head is fiber coupled.
Specifications
|
Receiver |
AIR-1550-TWM |
| Probe Laser |
Continuous-wave,
single-frequency fiber laser at 1550 nm
Power: 400mW-2W |
Generation
Laser |
Q-switched Nd:YAG at 1064
nm
Pulse width: 10 ns; pulse energy: 50, 200 or
400 mJ
Includes goggles and selection of mirrors
and lenses for directing and focusing the
beam |
Scanning
System |
Two linear stages and
controller; range of specifications
available |
Data
Acquisition
and Control |
LaserScan™ software:
Scanning system motion control, data
acquisition, processing and display
A-scan, B-scan, C-scan; specialized
processing
Desktop computer with DAQ card: PC running
Windows XP Pro
A/D board: 100 MS/s |
| Installation |
Included with system |
| Training |
IOS offers a fee-based
two-day seminar at IOS or customer site that
explains the principles of laser ultrasonics
and describes the operation of the LUKS
system |
| Customization |
The system components
described above can be modified to meet the
needs of the user The system components
described above can be modified to meet the
needs of the user The system components
described above can be modified to meet the
needs of the user |
| Pricing |
Please contact us for
pricing information should you be interested
in one of our standard systems or a
customized solution |
| Contact |
Dr. Marvin Klein, Manager,
Laser Ultrasonics Products Group
mklein@intopsys.com or +1(424) 263-6361 |
LaserScan™ Data Analysis Screen